5735 GW, Aarle-Rixtel  - The Netherlands  | +31492745710 | sales@mswil.com

Blog Post

What you need to know about the ABIRD (Active Background Ion Reduction Device)

What you need to know about the ABIRD (Active Background Ion Reduction Device)
ABIRD (Active Background Ion Reduction Device) Article Header

ABIRD (or Active Background Ion Reduction Device) is a popular accessory for nanospray-powered mass spectrometers, and has been approved for use on all Thermo Scientific instruments and nanospray sources.

ABIRD will consistently:

  • Isolate the instrument from spikes in lab air, which is critical for extended quantitative or comparative analyses.
  • Reduce background ion signals in full MS during nanospray-ESI analysis, enhancing S/N ratio quite dramatically in many cases.

These are benefits that only an ABIRD can provide, making it a smart enhancement for any nanospray ESI/LC-MS system.

Many users hope to demonstrate improved peptide detection with ABIRD installed, but get frustrated when they cannot easily do so. Given the improved signal-to-noise ratio ABIRD provides to a system, it might seem easy to find more peptides with ABIRD than without. We have found, however, that this is far from a simple task, and that there are many factors that need to be optimized in order to be able to demonstrate this in your own laboratory.

This blogpost outlines some of these factors and how you can further optimize them to enhance peptide detection with ABIRD in use. However, we can in no way guarantee this effect will be reproducible in your laboratory as there are simply far too many factors that might make that the entire LC-MS system operates sub-optimally and therefore not be able to demonstrate improved peptide detection with ABIRD in use.

The first important point is the correct installation of the ABIRD. This is usually a simple task of installing the device into position the outlet tube close to the instrument inlet. The Teflon tubing is resistant to the inlet temperatures, and it is fine to position it close to or even touch the heated inlet surface. This will fully flood the inlet with cleaned air and reduce background ion signals. You can install it in a variety of ways, and custom source installations are rarely a problem with flexible installation possibilities.

Figure 1. Three images of the ABIRD (separate or on a (Thermo) Source).

Next, the position of the spray tip needs to be optimized, as the gentle but consistent outflow of the ABIRD needs to be compensated for. This can best be done by infusing a tune solution via the nanospray tip at normal nanospray flow rates and spray voltages while tip position is adjusted. Generally, a low percentage of the spray plume is sampled, and placing the tip too close to the inlet reduces drying time of the droplets and thus reduces signal.

With ABIRD installed, observing the reduction in background ion signal is as simple as allowing the LC to reach initial equilibration (low % organic) then turning ABIRD on and off as you observe the signal in full scan mode across a wide mass range (see figure 2). Take note of the background ion noise level with ABIRD off, this is the approximate signal intensity level you will need to reach during these experiments. Introducing a very intense signal well above this level will not show more identified peptides by the use of ABIRD.

Figure 2. Visual representation of the differences in backbround ion signals with the ABIRD off (left) and on (right).

After this, you can begin to design your experiments to optimize data acquisition, LC conditions, instrument settings, and data analysis parameters to demonstrate improved peptide detection system-wide. Some of the parameters you must investigate include, but are not limited to:

  1. Instrument method parameters must be optimal. Minimum target values, maximum inject times, Top 20 vs Top 30 settings, dynamic exclusion settings, collision energies, fragmentation modes, and excluded charge states all need to be examined. A high sensitivity method will have several changes from a normal ID method, each instrument specific.
  2. If you are using the same method with Lock Mass (esp. Orbitrap) on, then you are wasting duty cycle looking for your low-level lock mass when ABIRD is on. So making a method that saves that duty cycle can help find more peptides.
  3. If you are DB searching only strict enzyme settings, you would not identify the vast number of partial cleavages that really exist at low levels.
  4. If your spray tip is not spraying consistently across the gradient, you will not find the low-level peptides it is looking for. Stable consistent spray is necessary for high sensitivity analysis.
  5. Chromatography is KEY! If you are running a very complex mixture on a short gradient with an overloaded column you will certainly miss the low-level peptides. We have found that most labs have a vast amount of improvement they can do on their chromatography systems and it can often make a significant difference in the overall laboratory operations.
  6. Your nanoLC MS system may leave room for improvement (tune and cal) or old multipliers/ MCP, then you will not gain much. There are numerous parameters that could be off on the MS instrument itself preventing you from seeing a benefit on Protein IDs.

Running daily standards for chromatography and peptide sensitivity to evaluate your instrument is mandatory. We have standard mixes available and are happy to supply them so we can do a real sensitivity check on your instrument as it stands and after optimizations. (Interested? Please contact us for the details.)

Think about it like this. You need to present the instrument more unique features/scans than it can isolate and analyze with the background ions present. When the extra background ions are reduced, the instrument has an extra duty cycle to find the lower-level peptides you are looking for. Also, without the high-level background ions, it will fill to normal capacity with whatever is left, and it will actually see peptides it did not see previously. This is a balance between how bad the background ions are and how well your instrument is optimized for the reduced background (ions) situation. All this is 100% user dependent/instrument dependent/problem dependent/data analysis dependent and obviously ABIRD cannot control for all this.

The newer & faster the instrument is, the more difficult it will be to demonstrate it. An ABIRD on an old LTQ or LTQ-Orbitrap will make a much more significant impact than the new fast instruments.

Keep in mind that this is an effect only observed in very low-level complex peptide mixtures. Peptide signals that approach background ion level signals will be improved most.


Please review this document and provide feedback and comments on it as you see fit (by mail: sales@mswil.com). We are happy to discuss these topics at length, and we might request a review of your raw data file to help this discussion. We look forward to your feedback and hope to make a difference in your daily lab operations.